MHRA software and AI as a medical device

MHRA regulation of software and AI as a medical device (Change Programme, AI Airlock and post-market surveillance rules)

Phasing inBindingRegulation

Software and AI with a medical purpose is regulated as a medical device in Great Britain. AI-specific rules are planned, not yet in force.

Read the official text (opens in a new tab)

Quick review

Key points

  • The rules are binding on manufacturers of software and AI that qualify as medical devices on the Great Britain market. The MHRA says many such products are regulated this way.
  • Post-market surveillance rules in force since 16 June 2025 require real-world safety data to be collected and serious incidents to be reported within 15 days. The limit is 2 days for a serious public health threat and 10 days for a death or an unanticipated serious deterioration in a person's health (regulation 44ZH(4) and (5)).
  • AI-specific changes are planned, not yet law. On 6 October 2026 the government accepted all 44 recommendations of the National Commission.
  • MHRA's AI Airlock sandbox opened Phase 3 applications on 6 October 2026, focused on post-market surveillance and lifecycle regulation.
  • No penalty figure was confirmed from the official text for this synopsis.

Version and dates

Version described
Medical Devices Regulations 2002 with the post-market surveillance amendment regulations (SI 2024/1368); AI-specific changes are planned and not yet law
Legal effect from
16 June 2025: Post-market surveillance rules took effect; the Medical Devices Regulations 2002 already applied
Latest change recorded
16 December 2024: Post-market surveillance amendment regulations (SI 2024/1368) became law; later AI-specific changes are drafts or plans only
Next date
December 2026: MHRA committed to issue draft guidance on managing changes to AI-enabled devices by December 2026.
Checked
16 November 2026 against the official source, and independently cross-checked. Every date and reference on this page was then audited against the official text; any that could not be confirmed says so.
This synopsis
Version 1.0, 16 November 2026 (first published)

Laws and standards change. Anything after the date checked is not reflected: confirm the current version in the official text.

Research starting point. For information only: a research starting point in plain English, not a definitive or complete guide and not legal advice. Validate it against the official text, challenge its conclusions where appropriate, and take advice from a qualified lawyer before acting.

1What it is

The Medicines and Healthcare products Regulatory Agency (MHRA) regulates medical devices, including software and AI that meet the definition of a medical device. Today this runs through the existing Medical Devices Regulations 2002, plus post-market monitoring rules in force since June 2025. The MHRA has a change programme, a testing sandbox called AI Airlock, and government-backed plans for AI-specific reform.

Approach: Sector-specific, meaning rules for one sector, such as banking or health.

2Does it apply to us?

Who is in scope
Manufacturers of software and AI that qualify as medical devices (or in vitro diagnostic devices) placed on the Great Britain market. Many products are regulated this way according to the MHRA. Whether a product qualifies depends on its intended purpose.
If you are based elsewhere
The post-market rules apply to devices placed on the market or put into service in Great Britain (England, Scotland and Wales). A non-UK business selling a device into Great Britain is affected as a manufacturer.

3Status and key dates

As checked on 16 Nov 2026. Dates can move: check the official text.

Phasing in. The Medical Devices Regulations 2002 and the post-market surveillance rules (in force 16 June 2025, the MHRA's date; legislation.gov.uk shows them as inserted into the 2002 Regulations on 17 June 2025) apply now; AI-specific changes are planned and not yet law, as of the sources read on 16 November 2026.

  1. 17 Oct 2022MHRA published the Software and AI as a Medical Device Change Programme Roadmap.
  2. 16 Dec 2024The post-market surveillance amendment regulations (SI 2024/1368) became law.
  3. 16 Jun 2025The post-market surveillance rules took effect for devices placed on the GB market on or after this date (regulation 44ZD(1) of the Medical Devices Regulations 2002).
  4. 8 Apr 2026MHRA announced GBP 1.2 million a year for 2026 to 2029 for the AI Airlock sandbox.
  5. 8 May 2026MHRA published draft pre-market regulations (draft Medical Devices (Amendment) Regulations 2026) on the WTO portal; survey closed 19 June 2026.
  6. 10 Sep 2026The National Commission into the Regulation of AI in Healthcare published its recommendations.
  7. 6 Oct 2026Government accepted all 44 Commission recommendations; AI Airlock Phase 3 applications opened.
  8. Dec 2026Still to comeMHRA committed to issue draft guidance on managing changes to AI-enabled devices by December 2026.
  9. 2027MHRA says it will begin consulting in 2027 on how to qualify and classify AI-enabled devices. A full implementation roadmap for all 44 Commission recommendations is due by Spring 2027.

4What it requires

The main obligations, in plain words. The official text has the full detail.

  • Manufacturers must collect and assess real-world safety and performance data for devices already in use (post-market surveillance rules in force 16 June 2025, described in the MHRA press release of 16 June 2025).
  • Serious incidents must be reported to the MHRA within 15 days (previously 30), or within 2 days for a serious public health threat and 10 days for a death or an unanticipated serious deterioration in a person's health (regulation 44ZH(4) and (5) of the Medical Devices Regulations 2002); Field Safety Notices go to the MHRA for review before they are shared with users (MHRA press release, 16 June 2025).
  • Manufacturers must provide post-market surveillance reports within 3 working days of a request (regulation 44ZF(5) of the Medical Devices Regulations 2002; the MHRA press release of 16 June 2025 says 3 days).
  • Software on its own can be a medical device. The MHRA publishes guidance on when software is a device and on reporting incidents (MHRA programme page).
  • Planned, not yet in force: the government has accepted proposals for lifecycle-long monitoring of AI devices, updated rules on what counts as an AI medical device, and a change-control approach for adaptive AI (Government response, 6 October 2026).

5Reporting and record-keeping

What has to be reported, notified, published, labelled, recorded or audited, and to whom.

  • Manufacturers must report serious incidents to the MHRA within 15 days (previously 30), including serious incidents relating to side effects. The limit is 2 days for a serious public health threat and 10 days for a death or an unanticipated serious deterioration in a person's health (regulation 44ZH(4) and (5) of the Medical Devices Regulations 2002).
  • Manufacturers must submit Field Safety Notices, their safety communications to users, to the MHRA for review before sharing them.
  • Manufacturers must provide post-market surveillance reports or Periodic Safety Update Reports within 3 working days of an MHRA request.
  • Manufacturers must regularly summarise and assess device performance over time under the new trend and summary reporting rules.

6Risks and penalties

What an organisation is exposed to under this instrument, as its text states.

  • For higher-risk devices, UK approved bodies monitor manufacturers' surveillance reports, so these products receive a higher level of scrutiny.
  • Planned, not yet in force: the government accepted a recommendation that the MHRA strengthen enforcement for software and AI devices, including using the civil sanctions regime in the Medicines and Medical Devices Act to issue financial penalties where appropriate. Details are due by Spring 2027.
  • The government response commits to assessing AI devices throughout their working life rather than relying on a one-time assessment, so manufacturers of AI devices face closer lifecycle monitoring once the reforms are in place.
Penalties
Not stated in the pages read. The post-market surveillance regulations amend the Medical Devices Regulations 2002; no penalty figure was confirmed from the official text for this synopsis.
Who enforces it
Medicines and Healthcare products Regulatory Agency (MHRA); Department of Health and Social Care (sponsor of the AI Airlock sandbox)

7What it could mean for an organisation

General information, not advice. Whether and how it applies depends on your facts: take legal advice.

  • If your organisation builds or sells software or AI for a medical purpose in Great Britain, the MHRA says many such products are regulated as medical devices.
  • If you are a manufacturer, the post-market surveillance rules in force since 16 June 2025 set 15-day serious incident reporting and a requirement to collect real-world performance data. The reporting limit is shorter in two cases: 2 days for a serious public health threat and 10 days for a death or an unanticipated serious deterioration in a person's health (regulation 44ZH(4) and (5)).
  • The AI-specific reforms are still plans: draft guidance on AI change control is due by December 2026, and a consultation on classification by Spring 2027. The draft pre-market regulations were out for comment in 2026 and are not in force.
  • AI Airlock is the MHRA's regulatory sandbox, where innovators, regulators and healthcare partners explore regulatory challenges using real products; Phase 3 applications opened on 6 October 2026 (MHRA press release of that date).

8Where to get help

Official guidance, tools and contacts for applying it. Links open the issuing body's own pages.

9Where it touches the SEAM governance points

The same six points MarLuc uses to read governance in every case study.

A1 Safe use
Device rules aim to ensure software and AI used for clinical purposes is safe and evidenced before and after it reaches patients.
A3 Monitoring and assurance
Post-market surveillance rules and AI Airlock Phase 3 focus on monitoring performance after deployment.
A4 Vendor AI
Manufacturers are the regulated party, so a health organisation buying AI software relies on the manufacturer's compliance.
A6 Standards and regulation
Draft guidance and reforms to 2027 show where the regime is heading for AI as a medical device.

Sources

  1. [1]MHRA, GOV.UK: Software and artificial intelligence (AI) as a medical device (opens in a new tab)Official source · gov.uk
  2. [2]MHRA, GOV.UK: Government backs recommendations of NHS doctors-led AI Commission (6 October 2026) (opens in a new tab)Official source · gov.uk
  3. [3]MHRA, GOV.UK: First major overhaul of medical device regulation comes into force across Great Britain (16 June 2025) (opens in a new tab)Official source · gov.uk
  4. [4]MHRA, GOV.UK: MHRA invites views on proposed changes to medical device regulation (11 May 2026) (opens in a new tab)Official source · gov.uk
  5. [5]MHRA, GOV.UK: MHRA expands AI Airlock programme with a GBP 3.6 million funding boost (8 April 2026) (opens in a new tab)Official source · gov.uk
  6. [6]MHRA, GOV.UK: Implementation of the future regulations of medical devices (opens in a new tab)Official source · gov.uk

Checked 16 Nov 2026 · independently cross-checked 16 Nov 2026 · dates and references audited 16 Nov 2026 · synopsis version 1.0

Spotted an error or a change? Tell us

Where does your organisation stand?

The free SEAM Check takes ten minutes. It shows where AI value is leaking in your organisation, including whether governance lets successful AI scale safely, with proportionate controls and managed regulatory exposure. It is a starting point, not legal advice.