United States (federal) · Health and medical devices · MarLuc synopsis
FDA AI-enabled device guidance
FDA guidance on AI-enabled device software functions: Predetermined Change Control Plans (final) and Lifecycle Management (draft)
In forceVoluntaryRegulator guidanceUS device regulator recommendations on pre-agreeing how an AI device may change after approval, and on managing AI devices over their life.
Read the official text (opens in a new tab)Quick review
Key points
- Both documents are non-binding FDA guidance; the lifecycle document is still a draft (January 2025) and the PCCP document is final (reissued August 2025).
- They apply to makers of medical devices with AI-enabled software functions that seek FDA marketing authorisation.
- A PCCP (predetermined change control plan) lets agreed changes proceed without a new marketing submission, if consistent with the authorised plan.
- A PCCP should have a Description of Modifications, a Modification Protocol and an Impact Assessment.
- The draft recommends a postmarket performance monitoring plan, as AI device performance may degrade.
Version and dates
- Version described
- PCCP final guidance (issued 4 December 2024, reissued 18 August 2025) and Lifecycle Management draft guidance (7 January 2025), still listed as draft at 9 April 2026
- First issued
- 4 December 2024: First issued (PCCP final guidance; Lifecycle Management draft issued 7 January 2025)
- Latest change recorded
- 18 August 2025: PCCP final guidance reissued
- Checked
- 16 November 2026 against the official source, and independently cross-checked. Every date and reference on this page was then audited against the official text; any that could not be confirmed says so. Some points could not be confirmed: see the note at the foot of the page.
- This synopsis
- Version 1.0, 16 November 2026 (first published)
Laws and standards change. Anything after the date checked is not reflected: confirm the current version in the official text.
Research starting point. For information only: a research starting point in plain English, not a definitive or complete guide and not legal advice. Validate it against the official text, challenge its conclusions where appropriate, and take advice from a qualified lawyer before acting.
1What it is
Two guidances from the US Food and Drug Administration for devices that include AI. The final one says what to put in a Predetermined Change Control Plan, a plan agreed in advance for how an AI device may be updated without a fresh submission each time. The draft covers lifecycle management and what to include in a marketing submission, including transparency, bias and monitoring.
Approach: Sector-specific, meaning rules for one sector, such as banking or health.
2Does it apply to us?
- Who is in scope
- Manufacturers of medical devices that include AI-enabled device software functions and seek FDA marketing authorisation (premarket approval, 510(k) clearance or De Novo).
- If you are based elsewhere
- FDA guidance relates to devices marketed in the United States. A manufacturer based outside the United States is relevant if it markets such a device in the US.
3Status and key dates
As checked on 16 Nov 2026. Dates can move: check the official text.
In force. The Predetermined Change Control Plan guidance is final (issued 4 December 2024, reissued 18 August 2025); the Lifecycle Management guidance (7 January 2025) was still listed as draft in an FDA list current as of 9 April 2026, and MarLuc could not confirm a later change.
- 4 Dec 2024PCCP final guidance first issued
- 7 Jan 2025Lifecycle Management draft guidance issued for comment (docket FDA-2024-D-4488)
- 18 Aug 2025PCCP final guidance reissued
4What it requires
The main obligations, in plain words. The official text has the full detail.
- Non-binding: FDA says the guidance does not establish rights and is not binding on FDA or the public (both documents).
- A PCCP should consist of a Description of Modifications, a Modification Protocol and an Impact Assessment (PCCP guidance, Introduction).
- A PCCP covers modifications that would generally otherwise need a new marketing submission; section 515C of the FD&C Act lets changes consistent with an authorised PCCP proceed without one (PCCP guidance).
- A PCCP may be established through the PMA, 510(k) or De Novo pathway (PCCP guidance).
- The draft lifecycle guidance recommends a total product life cycle approach with transparency and bias strategies, including evidence on demographic groups (draft guidance).
- The draft says manufacturers should have a postmarket performance monitoring plan, as AI device performance may degrade over time (draft guidance, section XI).
5Reporting and record-keeping
What has to be reported, notified, published, labelled, recorded or audited, and to whom.
- A PCCP is reviewed by the FDA as part of a marketing submission for the device; the guidance recommends what information the submission should include (PCCP guidance). Owed to the FDA.
- The draft lifecycle guidance recommends information on performance monitoring after the device is on the market (draft guidance, section XI). Owed to the FDA in the submission.
6Risks and penalties
What an organisation is exposed to under this instrument, as its text states.
- Under the PCCP guidance, deviations from an authorised PCCP would generally cause the device to be adulterated and misbranded under sections 501(f)(1) and 502(o) of the FD&C Act.
- The guidance says the FDA may take legal or regulatory action, including seizure or injunction, against prohibited acts involving adulterated or misbranded devices.
- The guidance itself is not binding and does not establish rights; an alternative approach is allowed if it meets the law.
- Penalties
- None stated in the guidance; it is non-binding. Medical device law itself carries enforcement powers, which MarLuc did not review for this entry.
- Who enforces it
- US Food and Drug Administration (Center for Devices and Radiological Health and other centres)
7What it could mean for an organisation
General information, not advice. Whether and how it applies depends on your facts: take legal advice.
- If you buy AI-enabled medical devices for US use, the device's authorised PCCP (if any) sets what changes the maker may make without a new FDA submission.
- If you make such devices, the final PCCP guidance lists three components to describe and justify planned changes.
- The lifecycle guidance is a draft and may change; it recommends transparency and bias evidence that buyers may ask suppliers about.
- FDA guidance is non-binding, but the statute (section 515C) and device regulations still apply.
8Where to get help
Official guidance, tools and contacts for applying it. Links open the issuing body's own pages.
- FDA: PCCP final guidance for AI-enabled device software functions (opens in a new tab)Full text of the final guidance.fda.gov
- FDA: Lifecycle Management draft guidance page (opens in a new tab)Draft guidance, docket and comment details.fda.gov
- FDA: Guidances with Digital Health Content (opens in a new tab)List of FDA digital health guidance documents.fda.gov
9Where it touches the SEAM governance points
The same six points MarLuc uses to read governance in every case study.
- A1 Safe use
- Pre-agreed change plans and validation set safe limits on how an AI device may change.
- A3 Monitoring and assurance
- The draft guidance recommends a postmarket performance monitoring plan.
- A4 Vendor AI
- Manufacturers of AI devices are the suppliers whose documentation a healthcare buyer relies on.
- A6 Standards and regulation
- A sector regulator's approach to AI change control and lifecycle management.
Sources
- [1]FDA: Marketing Submission Recommendations for a PCCP for AI-Enabled Device Software Functions (final guidance, reissued 18 August 2025) (opens in a new tab)Official source · fda.gov
- [2]FDA: AI-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations (draft guidance, 7 January 2025) (opens in a new tab)Official source · fda.gov
- [3]FDA: Lifecycle Management draft guidance web page (opens in a new tab)Official source · fda.gov
- [4]FDA: Guidances with Digital Health Content (current as of 9 April 2026) (opens in a new tab)Official source · fda.gov
Checked 16 Nov 2026 · independently cross-checked 16 Nov 2026 · dates and references audited 16 Nov 2026 · synopsis version 1.0
The cross-check could not confirm every point on this page from an official source that could be opened. Treat anything marked "could not confirm" as unverified, and check the official text.
Spotted an error or a change? Tell usWhere does your organisation stand?